Instructions for Canine Distemper Virus (CDV) Elisa Kit

Instructions for Canine Distemper Virus (CDV) Elisa Kit
This kit is for research use only.
Detection range: 96T
2ng / L-50ng / L
purpose of usage:
This kit is used to determine the content of CDV in canine serum, plasma and related liquid samples.
Experimental principle
Canine Distemper Virus (CDV) Elisa Kit Instructions This kit uses the double antibody sandwich method to determine the level of canine distemper virus (CDV) in the specimen. Canine Distemper
The toxic antibody is coated on the microwell plate to make a solid phase antibody, and then the distemper virus (CDV) is added to the microwell coated with the monoclonal antibody in turn, and then
The HRP-labeled distemper virus antibody binds to form an antibody-antigen-enzyme-labeled antibody complex, which is added after thorough washing
TMB color development. TMB is converted into blue under the catalysis of HRP enzyme, and into the final yellow under the action of acid. Yan
The color depth is positively correlated with the distemper virus (CDV) in the sample. Measure the absorbance (OD) at 450nm with a microplate reader
Value), calculate the concentration of canine distemper virus (CDV) in the sample by the standard curve.
Kit composition
1 30 times concentrated washing solution 20ml × 1 bottle 7 stop solution 6ml × 1 bottle
2 Enzyme label reagent 6ml × 1 bottle 8 standard (96ng / L) 0.5ml × 1 bottle
3 Enzyme label coated plate 12 wells × 8 strips 9 standard dilutions 1.5ml × 1 bottle
4 Sample diluent 6ml × 1 bottle 10 instructions 1 copy
5 Developer A solution 6ml × 1 bottle 11 2 sealing film
6 Developer B liquid 6ml × 1 / bottle 12 1 sealed bag
Specimen requirements
1. Specimens are extracted as soon as possible after collection, and extraction is performed according to relevant literature, and experiments should be conducted as soon as possible after extraction. If not
Perform the test immediately. The specimen can be stored at -20 ℃, but repeated freezing and thawing should be avoided
2. Samples containing NaN3 cannot be detected because NaN3 inhibits horseradish peroxidase (HRP) activity.
Steps
1. Dilution of standard products: this kit provides one original standard product, the user can dilute in a small test tube according to the following chart
release.
48ng / L No. 5 standard 150μl original standard added 150μl standard dilution
24ng / L No. 4 standard 150μl No. 5 standard added 150μl standard dilution
12ng / L No. 3 standard 150μl No. 4 standard added 150μl standard dilution
6ng / L No. 2 standard 150μl No. 3 standard added 150μl standard dilution
3ng / L No. 1 standard 150μl No. 2 standard added 150μl standard dilution
2. Add sample: set up blank wells separately (the blank control wells do not add samples and enzyme label reagents, the rest of the steps are the same), standard wells,
Sample hole to be tested. Add 50μl of the standard on the enzyme-coated plate accurately, and add 40μl of sample diluent to the sample well.
Then add 10μl of the sample to be tested (the final dilution of the sample is 5 times). Add the sample to the bottom of the well of the microtiter plate
Do not touch the wall of the hole, shake gently to mix.
3. Incubation: seal the plate with the sealing film and incubate at 37 ° C for 30 minutes.
4. Mixing solution: Dilute 30 times concentrated washing liquid with distilled water 30 times and reserve
5. Washing: Carefully peel off the sealing film, discard the liquid, spin dry, fill each well with washing liquid, let stand for 30 seconds and then discard, so
Repeat 5 times and pat dry.
6. Add enzyme: add 50μl of enzyme label reagent to each well, except blank well.
7. Incubation: The operation is the same as 3.
8. Washing: The operation is the same as 5.
9. Color development: add 50μl of developer A to each well, and then add 50μl of developer B, mix gently, and develop color at 37 ℃ in the dark
15 minutes.
10. Termination: Add 50μl of stop solution to each well to stop the reaction (at this time the blue will turn to yellow).
11. Determination: Measure the absorbance (OD value) of each well in sequence with the blank air conditioner at zero and 450nm wavelength. Determination should be terminated
Within 15 minutes after the solution.
Summary of operating procedures:
Calculation
Taking the concentration of the standard as the abscissa and the OD value as the ordinate, draw a standard curve on the coordinate paper, according to the sample
The corresponding concentration of OD value is found by the standard curve; multiplied by the dilution factor; or the standard concentration and OD value are used to calculate
The linear regression equation of the quasi-curve, the OD value of the sample is substituted into the equation, the sample concentration is calculated, and then multiplied by the dilution factor,
This is the actual concentration of the sample.
Precautions
1. The kit should be equilibrated at room temperature for 15-30 minutes before being taken out of the refrigerated environment.
After use, the slats should be stored in sealed bags.
2. Crystals may be precipitated in the concentrated washing liquid, which can be heated and dissolved in a water bath during dilution, and the results will not be affected during washing.
3. The sampler should be used at each step of sample addition, and the accuracy should be regularly checked to avoid test errors. One sample loading time is best
Control within 5 minutes, if the number of specimens is large, it is recommended to use a volley gun to add samples.
4. Please make a standard curve at the same time of each measurement, it is best to make a double hole. If the content of the test substance in the specimen is too high (sample OD value
Is greater than the OD value of the first well of the standard well), please dilute it with a certain multiple (n times) of the sample diluent before measuring.
When calculating, please multiply the total dilution factor (× n × 5).
5. The sealing film is limited to one-time use to avoid cross-contamination.
6. Please keep the substrate away from light.
7. Strictly follow the instructions, and the test results must be determined by the microplate reader.
8. All samples, washing liquids and various wastes should be treated as infectious agents.
9. The components of different batches of this reagent shall not be mixed.
10. If there is any difference with the English manual, the English manual shall prevail.
Storage conditions and validity period
1. Kit storage :; 2-8 ℃.
2. Validity: 6 months Instructions for Canine Distemper Virus (CDV) Elisa Kit

The most fashionable Leather Handbags from He-kayla brand, China's brand He-kayla launched a new interpretation of the new Totes Handbags ,simple,streamlined design and the emergence of a more daring in the hasp at the classic inscribed "He-kayla" logo, no doubt the new season will be the necessary choice in the dress for men and women. Soft leather Womens Messenger Bag including fine beige, elegant purple and classic black to create a low-key luxurious femininity and circulated out of infinite charm.



We not only production own brand but also accept customized, OEM and ODM we can do it, if interested any fashion Leather Shoulder Handbags or make as your original sample & specific request, just feel free to contact us at any time, then will be pay highly attention and provide the quotation ASAP.

 

The high quality Briefcase Handbags and  Satchel Handbags  can succeed in the large market, while the perfect technical support, pre-sales and after-sales service team will be able to efficiently and quickly provide the best quality service to customers. Hope to become your best strategic cooperative partners! 

Handbags

Woven Tote Handbags,Woven Handbags,Canvas Tote Handbag,Cotton Handbags,Women Handbags,PU Leather Handbags

Kang Jin Apparel Co., Ltd. , http://www.classicsbags.com